The main faults of the electromagnetic flowmeter in use are sensor failure, instrument failure, incorrect installation, external electromagnetic interference, lightning strikes, etc. The electromagnetic flowmeter is especially important for the analysis of the cause of the fault, and the inspection of the electromagnetic flowmeter is strengthened to make it play a greater role.
1. Electromagnetic flowmeter no flow signal output
First check if there is any problem with the power supply, line connection, terminal block, etc. Check if the sensor is normal and check if the coil is disconnected or shorted. Electromagnetic flowmeters are usually caused by water in the meter well. The water is drained and dried in time, and the sensor probe is cleaned. The meter can resume normal operation. Otherwise, if the soaking time is long, the excitation coil of the sensor may be burned out and the meter cannot be used again.
2. Electromagnetic flowmeter lightning and external electromagnetic interference
Lightning strikes induce transient high voltages and surge currents in the line, and entering the meter can damage the meter. From the analysis of damaged components in lightning faults, the induced high voltage and surge currents that cause faults are mostly introduced from the control room power line, so it is necessary to recognize the importance of setting the control room instrument power line lightning protection facilities. Damage to the electromagnetic flowmeter is large, lightning protection measures should be strengthened, the grounding should be checked to be good, and the damage of the lightning to the instrument should be reduced.
For external electromagnetic interference, the electromagnetic sensor and its pipeline can be insulated and insulated. The cable is placed in the protection tube, and anti-interference measures are adopted to obtain good measurement results.
3. Zero instability
If the pipe is not filled with water or there are air bubbles in the water, and the sensor is not well grounded, it will cause zero instability. If there are air bubbles, add a getter upstream.
4. The output value is incorrect
Check the instrument constant, pipe diameter and set value are correct, adjust the converter zero point, check whether the water in the pipe is full, check whether the length of the straight pipe is not enough.
5. Electromagnetic flowmeter common fault
Faults caused by damage to the components of the instrument itself; failures caused by improper selection, improper installation, environmental interference, fluid characteristics, etc., such as display fluctuations, reduced accuracy, and even instrument damage. As long as you pay attention to such faults in your daily work and solve them well, the electromagnetic flowmeter must be able to play its due role.